Misurazione non invasiva della bilirubina in Neonatologia: limiti e potenzialitą

Giulia De Bernardo1, Rosa Teresa Scaramuzzo1, Francesco Gabbriellini1, Barbara Barni1, Ilaria Vannozzi1, Cristina Tuoni1, Francesca Dini1, Massimiliano Ciantelli1, Antonio Boldrini1,2
1UO Neonatologia, Azienda Ospedaliero-Universitaria Pisana, Pisa; 2Universitą di Pisa


indirizzo per corrispondenza:giuliadebernardo7@gmail.com



Non-invasive bilirubin measurement in neonatal care: limits and opportunities

Key words
Bilirubin, Neonatal jaundice, Hyperbilirubinemia, Bilirubinometer

Summary
Background - Besides traditional bilirubin measurement methods, in the past few years transcutaneous measurement has become popular in Neonatal Care Units, as it is easy to use and painless. However, transcutaneous bilirubinometers are currently limited to a screening method for hyperbilirubinemia and are not recommended in newborns exposed to phototherapy.
Objective - The objective of this study is to explore the accuracy of a transcutaneous bilirubinometer by examining its performance both in term and preterm infants for jaundice screening, as well as during phototherapy sessions.
Methods - Model JM-105 Dräger is the bilirubinometer tested in the present research. Transcutaneous and serum measurements were performed simultaneously and eventually analysed.
Results - The bilirubinometer was tested on 116 newborns (gestational age ≥ 26 weeks). 158 paired samples were obtained, 35 of which during phototherapy. Comparison of serum and transcutaneous measurements showed a correlation of 0.85 and 0.89, in term and in preterm infants respectively.
Conclusions - The bilirubinometer JM-105 can be considered a reliable screening tool. Serum confirmation should be obtained if transcutaneous bilirubin levels exceed 11 mg/dl. In neonates exposed to phototherapy, the skin surfaces that are usually unexposed to light, such as the back in case of overhead spotlights and the inguinal region covered by the diaper in case of whole body exposure, may save these patients from frequent blood sampling. However, given the limited number of premature infants and phototherapy patients involved in the present study, the results obtained in these settings should be considered preliminary, and may be eventually verified with a higher number of cases.


L'accesso č riservato agli abbonati alla rivista che si siano registrati.
Per accedere all'articolo in formato full text č necessario inserire username e password.

Vuoi citare questo contributo?
G. De Bernardo, R.T. Scaramuzzo, F. Gabbriellini, B. Barni, I. Vannozzi, C. Tuoni, F. Dini, M. Ciantelli, A. Boldrini. MISURAZIONE NON INVASIVA DELLA BILIRUBINA IN NEONATOLOGIA: LIMITI E POTENZIALITą. Medico e Bambino pagine elettroniche 2018; 21(3) https://www.medicoebambino.com/?id=RIC1803_10.html




La riproduzione senza autorizzazione č vietata. Le informazioni di tipo sanitario contenute in questo sito Web sono rivolte a personale medico specializzato e non possono in alcun modo intendersi come riferite al singolo e sostitutive dell'atto medico. Per i casi personali si invita sempre a consultare il proprio medico curante. I contenuti di queste pagine sono soggetti a verifica continua; tuttavia sono sempre possibili errori e/o omissioni. Medico e Bambino non č responsabile degli effetti derivanti dall'uso di queste informazioni.

Unauthorised copies are strictly forbidden. The medical information contained in the present web site is only addressed to specialized medical staff and cannot substitute any medical action. For personal cases we invite to consult one's GP. The contents of the pages are subject to continuous verifications; anyhow mistakes and/or omissions are always possible. Medico e Bambino is not liable for the effects deriving from an improper use of the information.